Thrombin generation is extremely sensitive to preheating conditions.
Author(s): De Smedt E, Hemker HC
Publication: J Thromb Haemost, 2011, Vol. 9, Page 233-4
PubMed ID: 21062415 PubMed Review Paper? No
Purpose of Paper
Conclusion of Paper
Studies
-
Study Purpose
The purpose of this study was to investigate the effects of preheating time on thrombin generation in plasma.
Summary of Findings:
The endogenous thrombin potential was overestimated by 50% when specimens were preheated in the instrument for 2 min. Preheating in the instrument for more than 2 minutes resulted in specimen evaporation and up to a 4 degree C difference in temperature between wells. When specimens were preheated for 5-15 min on the plate heater, thrombin generation became consistent.
Biospecimens
Preservative Types
- None (Fresh)
Diagnoses:
- Normal
Platform:
Analyte Technology Platform Protein Hematology/ auto analyzer Pre-analytical Factors:
Classification Pre-analytical Factor Value(s) Preaquisition Biomarker level 5 pM tissue factor
0.5 pM tissue factor
Hematology/ auto analyzer Specific Incubation temperature 37 degrees C
Hematology/ auto analyzer Specific Incubation time 2 min
5 min
10 min
15 min
Hematology/ auto analyzer Specific Heating method Covered on plate-heater
In fluorometer