Purpose of Paper
Conclusion of Paper
Studies
-
Study Purpose
The purpose of this study was to determine the effects of heparin and gel separator tubes on the detection of cardiac troponin I, myoglobin ad CK-MB in blood specimens from patients with acute coronary syndrome.
Summary of Findings:
Collection tube type and anticoagulant had no effects on cardiac troponin I. Decreased myoglobin and CK-MB were found in specimens collected in heparin tubes without a gel separator compared to specimens collected in gel separator tubes with no anticoagulant.
Biospecimens
Preservative Types
- None (Fresh)
Diagnoses:
- Coronary Artery Disease
Platform:
Analyte Technology Platform Protein Immunoassay Pre-analytical Factors:
Classification Pre-analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Gel separator
Plain
Biospecimen Acquisition Anticoagulant Heparin
None
Immunoassay Specific Targeted peptide/protein Cardiac troponin I
Myoglobin